Please connect the power cord before you start the operation.Press the main power switch at right lower corner of the frontpanel. The power switch lig
5.3 START cycleClose the door after loaded the instruments, press STARTkey. 0.5S later, electrical lock will work, with “KA”sounds. If the ele
Operation manualSEA Series Steam Sterilizers8Fig1The first way: fig1,fig2The second way:clockwise rotate the instrument rack by 90 based on the fir
SEA Series Steam SterilizersOperation manual96.Printer installation and USBDate:C.NC.S.TVp1 Vp2 Vp31Pp1 Pp2 Pp31H.S.T:Max Min AveH.E.T:D.S.TC.E.TResul
Operation manualSEA Series Steam Sterilizers107.System setupWhen the device on the ready to work state, press MENU/C key for 3 S will enter to system
With 2 sub-menus(fig 7-5). Press STEP and QUICK key to select language. Press OK key to save the setup and back to upper level of menu. Press MENU/
Operation manualSEA Series Steam Sterilizers12After a long time use, maybe some small impurity will deposit on the draining filter. It will affect vac
Clean the water tank every week by using medical disinfectant. First drain the distilled water out, using cotton fabric with medical disinfectant to s
Operation manualSEA Series Steam Sterilizers14After using the sterilizer for a period, the scale and impurity will leave at the surface of sealing cov
10.Transport & StoragePower the machine off, unplug the power cord and let the chamber cool down10.3 The autoclave should be transported and stor
On the serial number plate at back of the machine, you may find the unit's model number,serial and reference number. Please double check the numb
The manufacturer provides a warranty for repair at not change to the customer for two years after the invoicedate with presentation of the invoice. Fo
SEA Series Steam SterilizersOperation manual1713.Appendix.Form 1 Work mode chartForm 2 Error code chart12345678Item 9ErrorConditionsensor errorsuper
1-2 Pre-heating 2-3 Pre-vacuum 3-4 Sterilization 4-5 Exhausting 5-6 Drying 6-7 Balance period 1-7 ster
SEA Series Steam SterilizersOperation manual19Yun circuit connection chart N L (RY-B38)Tranformaer21v (input) (RY-B38)Transformer 10V1.5A (input
Vacuum pumpYun pipingDrainging water soleroid valvenornolly openSteam solenoid valveChamberUsed water tankExhaust solenoiel valvenornrally openDressar
SEA Series Steam SterilizersOperation manual2114.Trouble shooting1.During venting or drying, the thump sounds. It means that the Condensed water colle
15.Important information1.The autoclave must be installed on a leveled counter top.2.When operated at a higher elevation of more than 500 meters,the m
Heaviness!!!Smother!!!StabilityImportant information on Safety1.Please check the voltage according to the nameplate before plug the power cord, Don&ap
ContentsOperation manualSEA Series Steam Sterilizers1.Description & Usage2.Specifications ...
SEA Series Steam SterilizersOperation manualYUN-22L YUN-17L Steam sterilizersDoor handleWater tankWater tank connectionUsed water tank connectionUsed
Operation manualSEA Series Steam SterilizersCaution & WarningWater leakage connectionB type deviceGrounded correctlyFigures2.SpecificationsChamber
SEA Series Steam SterilizersOperation manualFig 3-1 3.InstallationThe Autoclave should be installed in a drafty place allowing minimum of 10 cm space
Operation manualSEA Series Steam SterilizersWe installed the draining filter before the device leaving factory. (fig 3-2). Clean the draining filter o
In case of faults, please contact your distributor or the manufacture immediately.4.Control panel1.LCD screen Figure 4-1Letter row 1 Working state, sh