Repair methods: Cause Details of cause Repairing way Power a. Power cord is not connected. b. Power voltage is not proper. Checking power plug
7. Machine Check: a. Check motor brush and PC board: please open cover then dismantle rotor and loosen 5 screws at the baseboard and take apart th
OPERATION & MAINTENANCE MANUAL FOR 6&(1 PRACTICAL TYPE CENTRIFUGE LTD. +DJDYLVKVW,VUDHO7HO)D[
INDEX (Ⅰ) SPECIFICATION---------------------------------------------------------------- 1 (Ⅱ) ROTORANDACCESSORY------------------------------------
(Ⅱ) ROTOR AND ACCESSORY 1. Centrifugal force calculation: Use its speed and radius to calculate force as following formula. RCF= (rpm
(Ⅲ) INSTALLATION 1. When you take out centrifuge from carton, please check if there is any scratch, press trace and damage on knob or not. If yes, p
(Ⅳ) OPERATION 1. Tube & Sample Setting (1) The tubes in use must be placed in even number such as 2, 4, 6, 8. Glass tube must be placed i
2. Door Lock & Cover (1) Please close cover tightly, otherwise rotor can not rotate, cover can not be opened during rotation. (2) Plug in
4. Operation Procedure: (1) Turn the speed to be “0” then plug in the centrifuge. (2) Press【Manual Safe Button】and cover can be opened. (3) Place samp
(Ⅴ) REPAIR 1. Checking before and after operation * Check to see if there is any material or water inside chamber and keep it clean. * Check